Parameters:
Overall Performance | Throughput | 800 Tests/hour,1200Tests/hour with ISE(Optional) |
Analysis Method | End-point,Fixed-time,Rate(Kinetic) | |
Certificates | CE/ISO9001,ISO14001,ISO13485 | |
Sample & Reagent System | Sample Positions | 150 Sample Positions |
Reagent Positions | 180 Reagent Positions | |
Sample Volume | 1-70μl | |
Reagent Volume | 210-350μl | |
Probe | Teflon coating, with anti-collision function, liquid level detection function | |
Probe Washing | Automatic washing interior and exterior | |
Reagent Cooling | Refrigerated tray with idependent switch | |
Reaction System | Temperature Control | Water bath,37±0.1℃,real-time monitoring |
Cuvettes | 160 reusable cuvettes, optical length 6mm | |
Mixer Probe | Independent stirring | |
Washing | Automatic cuvettes washing | |
STAT Function | YES | |
Optical System | Light Source | 12V/10W halogen lamp/fiber bragg grating optional |
Spectrophotometry | Post-spectral spectrophotometry | |
Wavelength | 12 wavelengths from 340-800nm | |
Absorbance | 0-3.5Abs | |
Calibration & QC | Calibration | Linear: K factor,1-point,2-point and multipoint linear Non-Linear:Spline,Polygon,Index,Ogarithm,Logit-4P,Logit-5P |
Quality Control | Real-time QC,Westgard multi rule,Cumulative sum check,Twin Plot(2D) | |
Data Management | Software | Windows 7/8/10,32 or 64 bit |
LIS System | Bi-direction,support HL7 protocol | |
Interface | LAN port access | |
Printer | External, multiple reporting mode avaliable | |
Working Conditions | Power Supply | AC220V±10%,50/60Hz,110 V±10%,60Hz,1500W |
Temperature | 15-30℃(±0.2℃) | |
Water Consumption | Deionized water:<25L/H | |
Humidity | 40%-85% | |
Size & Weight | External Size(W*D*H) | 1360*795*1155mm |
Net Weight | 290kg | |
Package Size(W*D*H) | 1718*958*1427mm | |
Gross Weight | 400kg |
Parameters:
Overall Performance | Throughput | 800 Tests/hour,1200Tests/hour with ISE(Optional) |
Analysis Method | End-point,Fixed-time,Rate(Kinetic) | |
Certificates | CE/ISO9001,ISO14001,ISO13485 | |
Sample & Reagent System | Sample Positions | 150 Sample Positions |
Reagent Positions | 180 Reagent Positions | |
Sample Volume | 1-70μl | |
Reagent Volume | 210-350μl | |
Probe | Teflon coating, with anti-collision function, liquid level detection function | |
Probe Washing | Automatic washing interior and exterior | |
Reagent Cooling | Refrigerated tray with idependent switch | |
Reaction System | Temperature Control | Water bath,37±0.1℃,real-time monitoring |
Cuvettes | 160 reusable cuvettes, optical length 6mm | |
Mixer Probe | Independent stirring | |
Washing | Automatic cuvettes washing | |
STAT Function | YES | |
Optical System | Light Source | 12V/10W halogen lamp/fiber bragg grating optional |
Spectrophotometry | Post-spectral spectrophotometry | |
Wavelength | 12 wavelengths from 340-800nm | |
Absorbance | 0-3.5Abs | |
Calibration & QC | Calibration | Linear: K factor,1-point,2-point and multipoint linear Non-Linear:Spline,Polygon,Index,Ogarithm,Logit-4P,Logit-5P |
Quality Control | Real-time QC,Westgard multi rule,Cumulative sum check,Twin Plot(2D) | |
Data Management | Software | Windows 7/8/10,32 or 64 bit |
LIS System | Bi-direction,support HL7 protocol | |
Interface | LAN port access | |
Printer | External, multiple reporting mode avaliable | |
Working Conditions | Power Supply | AC220V±10%,50/60Hz,110 V±10%,60Hz,1500W |
Temperature | 15-30℃(±0.2℃) | |
Water Consumption | Deionized water:<25L/H | |
Humidity | 40%-85% | |
Size & Weight | External Size(W*D*H) | 1360*795*1155mm |
Net Weight | 290kg | |
Package Size(W*D*H) | 1718*958*1427mm | |
Gross Weight | 400kg |